Main Characters
|
Description
|
Sample Gas Temperature
|
0 to 700℃, 700 to 1400℃(0.15 m length probe and High Temperature Probe Adapter ZO21P-H)
|
Sample Gas Pressure
|
5 to + 250kPa(compensated type recommended exceeds 5kPa)
|
Oxygen concentration of sample gas
|
not more than that found in normal air, typically 21vol%
|
Probe Length
|
0.15, 0.4, 0.7, 1.0, 1.5, 2.0m
|
Probe Material
|
SUS316 (JIS)
|
Ambient Temperature
|
-20 to +60℃(-20 to +150°C on the terminal box surface)
|
Reference Gas System
|
Instrument Air
|
Instrument Air System
|
50kPa + the pressure inside the furnace
|
Wetted Material
|
SUS316 (JIS), Zirconia, SUS304 (JIS) or ASTM grade 304 (flange), Hastelloy B, Inconel 600, 601)
|
Gas Connection:
|
Rc1/4 or 1/4 NPT (Female)
|
Wiring Connection
|
M20 × 1.5, 1/2 NPT
|
Installation
|
Flange mounting
|
Probe Mounting Angle
|
Horizontal to vertically downward.
|